Section 6 Instrument Simulator
6.23
Timed SRQs Tn
The Timed SRQs command,
Tn
, enables the Analyzer488 to generate a Service
Request (SRQ) after a specified amount of time has passed. The time period may be
between 0 and 65535 milliseconds (0 to 65 seconds) in increments of 1 millisecond.
SYNTAX:
T0
Disable SRQ on timeout (default)
Tn
Generate an SRQ at the end of
n
milliseconds (
n
=
0 to 65535)
T?
Return the amount of time (in milliseconds)
remaining before a timed SRQ will be generated
EXAMPLES:
PRINT#1,"CLEAR10"
reset the Analyzer488
PRINT#1,"OUTPUT10;T5000X"
generate an SRQ after 5 seconds
after 5 seconds, the Analyzer488
will generate an SRQ.
PRINT#1,"SPOLL10"
serial poll the Analyzer488
INPUT#2,C$
PRINT C$ display shows 64 since the
Analyzer488 was the source of the
Service Request.