Section 4 Serial Controller
4.65
RESET
The High Level Controller Mode command
RESET
provides a warm start of the
Analyzer488 interface. Issuing the
RESET
command clears the serial input (pending
commands) and output (pending data) buffers and reinitializes the internal IEEE
controller hardware. It is equivalent to issuing the following commands:
ABORT
DISARM
ERROR
OFF
LOCAL
TIME
OUT
10
The
RESET
command provides a warm start of the interface as well as clearing
all error conditions. Upon detection of the
RESET
command, the Analyzer488
unasserts its serial output handshake line (RTS). It reasserts RTS when it is capable of
accepting serial input data. If
XON
/
XOFF
handshake is selected, the handshake state is
not effected by the
RESET
command.
This command does not have a query option.
TYPE High Level Controller
SYNTAX
RESET
RESPONSE None
MODE High Level Controller
BUS STATES IFC,*IFC,*REN
QUERY None
RESPONSE
EXAMPLE PRINT#1,"RESET"