Measurement-computing Serial488A Manuel d'utilisateur Page 9

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 85
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 8
1.7
Serial488/512K
IEEE 488-1978
Implementation: C1, C2, C3, C4 and C28 controller subsets.(Serial to IEEE)
SH1, AH1, T6, TE0, L4, LE0, SR1, RL0, PP0, DC1, DT0, E1.
Terminators: Selectable CR, LF, LF-CR and CR-LF with EOI.
Connector: Standard IEEE 488 connector with metric studs.
Serial Interface.
EIA RS-232C: AB, BA, BB, CA, CB
Character Set: Asynchronous bit serial.
Output Voltage: 5 volts typical (RS-232C).
Input Voltage: ±3 volts min.; ±15v max.
Baud Rate: Selectable 110, 300, 600, 1200, 1800, 2400, 3600, 4800, 7200,
9600, and 19,200.
Data Format: Selectable 7 or 8 data bits; 1 or 2 stop bits; odd, even, mark,
space and no parity on transmit.
Duplex: Full with Echo/No Echo.
Serial Control: Selectable CTS/RTS or XON/XOFF.
Terminators: Selectable CR, LF, LF-CR and CR-LF.
Connector: 25-pin Sub-D male. DCE Configured.
General
Data Buffer: 512,000 characters.
Indicators: LEDs for IEEE Talk and Listen, Serial Send and Receive, and
Power.
Power: 105-125V or 210-250V; 50-60 Hz, 10 VA Max.
Dimensions: 188mm deep x 140mm wide x 68mm high (7.39" x 5.5" x
2.68").
Weight: 1.95 kg. (4.35 lbs).
Environment: 0° - 50°C; 0 to 70% R.H. to 35°C. Linearly derate
3% R.H./°C from 35° to 50°C.
Controls: Power Switch (external)
IEEE and Serial parameter switches (internal).
*Specifications subject to change without notice
.