2.2
2.2 Configuration
Three DIP switches internal to the
Serial488A
set the configuration of the
interface. NOTE: Selectable functions are read ONLY at power-on and should only
be set prior to applying power to the interface. The following figures illustrate the
factory default settings which are:
Serial Port: IEEE:
9600 Baud Mode = IEEE 488 Controller
8 Data Bits Address = 10
2 Stop Bits Bus Terminator = LF; EOI Disabled
No Parity Talk-back on Terminator Enabled
Serial Terminator = LF Talk-back on Time Out Enabled
Echo Disabled
RTS/CTS Handshake
SW3 Factory Default Settings
OPEN
12345678
SW3
IEEE Addr
IEEE Term
EOI
10
LF
Disabled
DOT
Switch
Side
View
Commentaires sur ces manuels