Measurement-computing CB-COM-7012 Manuel d'utilisateur Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 72
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
28
CB-7012, CB-7014, CB-7017 Users Manual
2.5 $AA0
Description: Perform Span Calibration
Syntax: $AA0[CHK](cr)
$ Delimiter character
AA Address of module to set (00 to FF)
0 Command for performing span calibration
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax or communication error may get no response.
! Delimiter for valid command
? Delimiter for invalid command
AA Address of responding module (00 to FF)
Example:
Command: $010 Receive: !01
Perform Address 01 span calibration, return successful.
Command: $020 Receive: ?02
Perform Address 02 span calibration; return calibration not
enabled before performing calibration command.
Related Command:
Sec. 2.6 $AA1, Sec. 2.18 ~AAEV
Related Topics:
Sec. 1.9 Calibration
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 71 72

Commentaires sur ces manuels

Pas de commentaire