Measurement-computing CB-7011 Manuel d'utilisateur Page 26

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 68
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 25
26
CB-7011/CB-7011D, CD-7018 Users Manual
2.5 $AA0
Description: Perform Span Calibration
Syntax: $AA0[CHK](cr)
$ delimiter character
AA address of setting module (00 to FF)
0 command for performing span calibration
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax error or communication error may get no re-
sponse.
! delimiter for valid command
? delimiter for invalid command or the calibration is not
enabled
AA address of response module(00 to FF)
Example:
Command: $010 Receive: !01
Perform address 01 span calibration, return success.
Command: $020 Receive: ?02
Perform address 02 span calibration, return the calibration is
not enabled before perform calibration command.
Related Command:
$AA1 (Sec. 2.6), ~AAEV (Sec. 2.20)
Related Topics:
Calibration (Sec. 1.9)
Vue de la page 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 67 68

Commentaires sur ces manuels

Pas de commentaire