USB-2416 User's Guide Specifications
28
Table 8. S type thermocouple accuracy specifications, including CJC measurement error.
All specifications are (±).
A/D data rate
Sensor
temperature range
Accuracy error
maximum
Accuracy error
typical
Tempco
(°C/°C)
3750 S/s
-50 °C 4.510 °C 2.930 °C
0.088
250 °C 2.165 °C 1.468 °C
1768 °C 3.187 °C 2.597 °C
2000 S/s
-50 °C 4.510 °C 2.930 °C
0.088
250 °C 2.165 °C 1.468 °C
1768 °C 3.187 °C 2.597 °C
1000 S/s
-50 °C 4.510 °C 2.930 °C
0.088
250 °C 2.165 °C 1.468 °C
1768 °C 3.187 °C 2.597 °C
500 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
100 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
60 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
50 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
25 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
10 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
5 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
2.5 S/s
-50 °C 3.798 °C 2.226 °C
0.088
250 °C 1.853 °C 1.156 °C
1768 °C 2.922 °C 2.332 °C
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.