USB-2416 User's Guide Specifications
27
Table 7. R type thermocouple accuracy specifications, including CJC measurement error.
All specifications are (±).
A/D data rate
Sensor
temperature range
Accuracy error
maximum
Accuracy error
typical
Tempco
(°C/°C)
3750 S/s
-50 °C 4.826 °C 3.133 °C
0.082
250 °C 2.117 °C 1.424 °C
1768 °C 2.842 °C 2.347 °C
2000 S/s
-50 °C 4.826 °C 3.133 °C
0.082
250 °C 2.117 °C 1.424 °C
1768 °C 2.842 °C 2.347 °C
1000 S/s
-50 °C 4.826 °C 3.133 °C
0.082
250 °C 2.117 °C 1.424 °C
1768 °C 2.842 °C 2.347 °C
500 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
100 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
60 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
50 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
25 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
10 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
5 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
2.5 S/s
-50 °C 4.065 °C 2.379 °C
0.082
250 °C 1.805 °C 1.113 °C
1768 °C 2.619 °C 2.123 °C
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after