Vue de la page 53
-4.5
-4
-3.5
-3
-2.5
-2
-1.5
-1
-0.5
0
0 25 50 75 100 125 150 175 200 225 250
F )IN (kHz
Gain (dB)
-0.3
-0.25
-0.2
-0.15
-0.1
-0.05
0
0 5 10 15 20 25 30 35 40 45 50
F
Gain (dB)
6-6 Analog Signals 878595 ZonicBook/618E
IN (kHz)
F
IN
kHz
F
IN
(kHz), Zoom-In to Filter Cutoff Frequency Region
F
IN
kHz
Filter settings are discussed in the separate software documentation.
DC Accuracy
The bypass filter scenario is most commonly used in applications where DC signal information is
important.
DC Accuracy, Excluding Noise
Applies to 2-Pole and Bypass Filter Modes
Accuracy at 0° to 50°C (32° to 122°F)
± % of Reading
Range
Typical Maximum
Offset
±25V
0.5 1.0 ±15mV
±5V
0.15 0.3 ±3mV
±2.5V
0.15 0.3 ±2mV
±1V
0.15 0.3 ±1mV
±500mV
0.15 0.3 ±1mV
±250mV
0.15 0.3 ±1mV
±100mV
0.15 0.4 ±1mV
±50mV
0.15 0.4 ±1mV
±25mV
0.2 0.4 ±1mV
These numbers are valid for 1 year after calibration and over the
entire operating temperature range of the unit.