Measurement-computing WBK Options Manuel d'utilisateur Page 56

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 216
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 55
CAUTION
Make the DB9 pin # 2 connection to the “low-side” or the “high-side” of the
circuit. BUT NEVER TO BOTH! Doing so will create a short circuit that could
damage equipment.
Internal Shunt Resistor, Low-Side Connection
Full Bridge (+) with Internal Shunt Resistor Installed at R1 on the WBK16/LC
DB9 Pin 2 Connected to the Low-Side
Internal Shunt Resistor, High-Side Connection
Full Bridge (+) with Internal Shunt Resistor Installed at R1 on the WBK16/LC
DB9 Pin 2 Connected to the High-Side
CAUTION
Make the DB9 pin # 2 connection to the “low-side” or the “high-side” of the
circuit. BUT NEVER TO BOTH! Doing so will create a short circuit that could
damage equipment.
Note 1: The switches represented in the schematics are controlled by software.
WBK16, pg. 16 949794 WBK16, Strain-Gage Module
Vue de la page 55