USB-5203 User's Guide Specifications
25
Thermistor measurement accuracy
The error shown in Table 7 below does not include errors of the sensor itself. The sensor linearization is
performed using a Steinhart-Hart linearization algorithm. These specs are for one year while operating the
device between 15 °C and 35 °C. The specification does not include lead resistance errors for 2-wire thermistor
connections. Please contact your sensor supplier for details on the actual sensor error limitations.
Total thermistor resistance on any given channel pair must not exceed 180 kΩ (refer to Table 8 below).
Table 7. Thermistor measurement accuracy specifications
Thermistor Temperature Range (°C) Maximum Accuracy Error (°C)
Ix+ = 10 µA
2252 Ω 40 to120 ±0.05
3000 Ω 40 to120 ±0.05
5000 Ω 35 to120 ±0.05
10000 Ω 25 to120 ±0.05
30000 Ω 10 to120 ±0.05
Typical resistance values at various temperatures for supported thermistors are shown in Table 8 below.
The device cannot measure resistance values greater than 180 k in thermistor mode. The 180 kΩ resistance
limit includes the total resistance across the current excitation (Ix+/Ix) pins, which is the sum of the thermistor
resistance and the lead resistances.
For accurate three-wire compensation, the individual lead resistances connected to the Ix+/Ixpins must be of
equal value.
Table 8. Typical thermistor resistance specifications
Temp
(°C)
2252 Ω
thermistor
3000 Ω thermistor 5 kΩ thermistor 10 kΩ thermistor 30 kΩ thermistor
40 76 kΩ 101 kΩ 168 kΩ 240 kΩ 885 kΩ
35 55 kΩ 73 kΩ 121 kΩ 179 kΩ 649 kΩ
30 40 kΩ 53 kΩ 88 kΩ 135 kΩ 481 kΩ
25 29 kΩ 39 kΩ 65 kΩ 103 kΩ 360 kΩ
20 22 kΩ 29 kΩ 49 kΩ 79 kΩ 271 kΩ
15 16 kΩ 22 kΩ 36 kΩ 61 kΩ 206 kΩ
10 12 kΩ 17 kΩ 28 kΩ 48 kΩ 158 kΩ
5 9.5 kΩ 13 kΩ 21 kΩ 37 kΩ 122 kΩ
0 7.4 kΩ 9.8 kΩ 16 kΩ 29 kΩ 95 kΩ
Throughput rate to PC
The analog inputs are configured to run continuously. Each channel is sampled twice per second. The maximum
latency between when a sample is acquired and the temperature data is provided by the USB unit is
approximately 0.5 seconds.
Maximum throughput to a CompactFlash® memory card is 1 S/s per channel.