Measurement-computing TempScan/1100 Manuel d'utilisateur Page 238

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 306
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 237
A-42 API Command Reference
889897
TempScan / MultiScan User’s Manual
L#
- Set Scan Rate
TempScan/1100 only
TYPE
System
EXECUTION
Deferred
SYNTAX
L#rate
Specify whether to select the optional slow scan rate or to select the default
fast scan. Valid options for
rate
are:
0
- Select fast scan rate (default),
and
1
- Select slow scan rate.
DESCRIPTION
Although the TempScan/1100 factory default setting is fast scan rate, the Set Scan Rate (
L#
) command can
program the scan rate to be reduced by a factor of 4 (from 16 readings per line cycle to 4 readings per line cycle).
This slow scan rate feature allows for increased settling time when measuring signals with long wire lengths.
Using this
L#
command, the fast scan rate can also be reinstated.
EXAMPLE
PRINT#1,"OUTPUT07;L#0X"
‘ Select fast scan rate
PRINT#1,"OUTPUT07;L#1X"
‘ Select slow scan rate
Vue de la page 237
1 2 ... 233 234 235 236 237 238 239 240 241 242 243 ... 305 306

Commentaires sur ces manuels

Pas de commentaire