Measurement-computing PCM-DAS16x/16 Manuel d'utilisateur Page 30

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 40
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 29
6.2.7 Isolated Grounds / Differential Inputs
Optimum performance with isolated signal sources is assured with the use of the dif-
ferential input setting. The diagram below shows the recommend connections is this
configuration..
26
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal Source
and A/D Board
Already Isolated.
Already isolated signal source
and A/D board connected to
a differential input.
GND
These grounds are
electrically isolated.
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your system’s noise immunity.
Vue de la page 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 39 40

Commentaires sur ces manuels

Pas de commentaire