Section 5 Command Descriptions
5.63
READ
The
READ
command is used to inspect the contents of a defined Macro Buffer.
Macro Buffers can be individually specified by their number. When a
READ
command
is received, a copy of the Macro buffer requested is sent to the serial output.
SYNTAX
READ[;][number
]
number
is a Macro Buffer number from
0
to
99
. If not specified,
Macro 0 is assumed.
RESPONSE The contents of the Macro buffer are returned to the serial output.
MODE Any
BUS STATES None
EXAMPLES
PRINT#1,"READ"
Read the contents of Macro 0
PRINT#1,"READ 20"
Read the contents of Macro 20