Measurement-computing Extender 488HS Rev.2.0 Manuel d'utilisateur Page 26

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 33
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 25
Section 3 IEEE 488 Primer
3.8
3.7.6 Local Lockout (LLO)
Issuing the
LLO
command prevents manual control of the instrument's
functions. ($11)
3.7.7 Device Clear (DCL)
This command causes all bus devices to be initialized to a
pre-defined or power up state. ($14)
3.7.8 Selected Device Clear (SDC)
This causes a single device to be initialized to a pre-defined or power up
state. ($04)
3.7.9 Serial Poll Disable (SPD)
The
SPD
command disables all devices from sending their Serial Poll
status byte. ($19)
3.7.10 Serial Poll Enable (SPE
)
A device which is Addressed to Talk will output its Serial Poll status
byte after
SPE
is sent and
ATN
is unasserted. ($18)
3.7.11 Group Execute Trigger (GET)
This command usually signals a group of devices to begin executing a
triggered action. This allows actions of different devices to begin
simultaneously. ($08)
3.7.12 Take Control (TCT)
This command passes bus control responsibilities from the current
Controller
to another device which has the ability to control. ($09)
Vue de la page 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 32 33

Commentaires sur ces manuels

Pas de commentaire