DBK65 8-Channel Transducer Interface Module The DBK65 is campatible with: WaveBook, ZonicBook, LogBook, DaqBook, DaqLab, DaqScan, and DaqBoard/20
Tip: Label each transducer with its associated channel/sub-channel information. 2. For a single DBK65 module, connect one end of the P1 cable to
DBK Option Cards and Modules 987693 DBK65 pg. 11 Calibrating a Transducer using the “Shunt Calibration Technique” The “shunt calibration” techn
DBK65 pg. 12 987693 DBK Option Cards and Modules DBK65 Specifications Dimensions: 285 mm W x 220 mm D x 45 mm H (11” x 8.5” x 1.75”) Weight: 1
DBK65 I-V CURVE AT 5V EXCITATION0.0001.0002.0003.0004.0005.0006.0000.12350.12000.11860.11640.10900.08970.07610.06620.05850.05250.04790.04370.04020.0
DBK65 pg. 14 987693 DBK Option Cards and Modules
Each of the 8 channels can be set for a different excitation voltage. 5, 10, 15, and 24 VDC are provided internally from the DBK65 and are selected v
Selecting an Excitation Voltage Each channel has a voltage select header, which consists of 5 pairs of pins and a jumper. The jumper position determ
Customizing a Voltage To make use of the custom voltage feature you will need to acquire a resistor of the calculated value. The formula to use is:
Channel PGM Resistor Location Voltage Out Jumper Header Voltage Set VoutResistor Value R2CH00 R110 J11 CH01 R120 J21 CH02 R130 J31 CH
5. Using rosin core solder and proper soldering technique, solder the resistor into position for the applicable channel. Be sure that the resistor
Configuring the DBK65 Address You can connect 1 or 2 DBK65 modules to a single main channel on the primary data acquisition device. Thus, a 16-chann
Configuring the Primary Data Acquisition Device DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration Use of a DBK65 with a DaqBo
Connecting the DBK65 to Signals and to the Primary Data Acquisition Device You can connect the DBK65 module to your primary data acquisition device a