DAC488 User’s Manual DAC488 Calibration 49
Part III: Automatic Calibration Report
After the automatic calibration finishes, a calibration report will be printed if the DAC488 is connected to
an IEEE printer. The following example shows the calibration report of the three volt ranges for Port 1:
DAC488/4 Performance Verification Port # 1
Nominal Actual Error Result Nominal Actual Error Result
10.0000 10.0000 0.0000 Pass -10.0000 -10.0000 0.0000 Pass
9.0000 8.9979 0.0021 Pass -9.0000 -8.9976 0.0024 Pass
8.0000 7.9997 0.0003 Pass -8.0000 -7.9994 0.0006 Pass
7.0000 7.0003 0.0003 Pass -7.0000 -7.0001 0.0001 Pass
6.0000 6.0011 0.0010 Pass -6.0000 -6.0009 0.0009 Pass
5.0000 4.9999 0.0001 Pass -5.0000 -4.9997 0.0003 Pass
4.0000 4.0004 0.0004 Pass -4.0000 -4.0002 0.0002 Pass
3.0000 2.9998 0.0003 Pass -3.0000 -2.9996 0.0004 Pass
2.0000 2.0005 0.0005 Pass -2.0000 -2.0004 0.0004 Pass
1.0000 0.9993 0.0007 Pass -1.0000 -0.9992 0.0008 Pass
0.0000 0.0000 0.0000 Pass -0.0000 -0.0000 0.0000 Pass
5.0000 5.0000 0.0000 Pass -5.0000 -5.0000 0.0000 Pass
4.5000 4.4989 0.0011 Pass -4.5000 -4.4988 0.0012 Pass
4.0000 3.9998 0.0003 Pass -4.0000 -3.9997 0.0003 Pass
3.5000 3.5002 0.0002 Pass -3.5000 -3.5001 0.0001 Pass
3.0000 3.0005 0.0005 Pass -3.0000 -3.0005 0.0005 Pass
2.5000 2.4999 0.0001 Pass -2.5000 -2.4999 0.0001 Pass
2.0000 2.0001 0.0001 Pass -2.0000 -2.0002 0.0002 Pass
1.5000 1.4999 0.0001 Pass -1.5000 -1.4998 0.0002 Pass
1.0000 1.0002 0.0002 Pass -1.0000 -1.0002 0.0002 Pass
0.5000 0.4996 0.0004 Pass -0.5000 -0.4996 0.0004 Pass
0.0000 0.0000 0.0000 Pass -0.0000 -0.0000 0.0000 Pass
1.0000 1.0001 0.0001 Pass -1.0000 -1.0000 0.0000 Pass
0.9000 0.8998 0.0002 Pass -0.9000 -0.8997 0.0003 Pass
0.8000 0.8000 0.0000 Pass -0.8000 -0.7999 0.0001 Pass
0.7000 0.7001 0.0001 Pass -0.7000 -0.7000 0.0000 Pass
0.6000 0.6001 0.0001 Pass -0.6000 -0.6000 0.0000 Pass
0.5000 0.5000 0.0000 Pass -0.5000 -0.4999 0.0001 Pass
0.4000 0.4001 0.0001 Pass -0.4000 -0.3999 0.0001 Pass
0.3000 0.3000 0.0000 Pass -0.3000 -0.2998 0.0002 Pass
0.2000 0.1995 0.0005 Pass -0.2000 -0.1993 0.0007 Pass
0.1000 0.0997 0.0003 Pass -0.1000 -0.0995 0.0005 Pass
0.0000 0.0000 0.0000 Pass -0.0000 -0.0000 0.0000 Pass
Automatic Calibration Report Format
A report similar to the one above will be printed for all four ports. A description of each item in the auto-
calibration report is given below:
• Nominal: The voltage the DAC was programmed to output.
• Actual: The actual voltage measured by the DMM.
• Error: The difference between the Nominal and the Actual values.
• Result: Indicates Pass if the port is functioning within the stated specifications.