DBK84 14-Channel Thermocouple Module
Overview …… 1
Hardware Setup …… 2
Module Connection …… 2
Open Thermocouple Detection …… 3
Module Configuration …… 4
DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration …… 5
DaqBook/2000 Series and DaqBoard/2000 Series Configuration …… 5
Software Setup …… 6
Using a Temperature Calibrator …… 7
DBK84 – Specifications …… 8
Reference Notes:
o Chapter 2 includes pinouts for P1, P2, P3, and P4. Refer to pinouts applicable to
your system, as needed.
o In regard to calculating system power requirements, refer to DBK Basics located near
the front of this manual.
Overview
The DBK84 is used in temperature measurement applications and provides connections for 14
thermocouples through convenient mini-TC connectors.
The DBK84 features on-board cold junction compensation (CJC) for direct measurement of type J, K, T, E,
N28, N14, S, R, and B thermocouples. The following table provides the temperature range for each of
these thermocouple types.
Thermocouple Temperature Ranges
T/C Type J K T E N28 N14 S R B
Temperature
Range °C
-200 to
760
-200 to
1200
-200 to
400
-270 to
650
-270 to
400
0 to
1300
-50 to
1768
-50 to
1768
50 to
1780
Temperature
Range °F
-328 to
1400
-328 to
2192
-328 to
752
-454 to
1202
-454 to
752
32 to
2372
-58 to
3214
-58 to
3214
122 to
3236
Up to fourteen external thermocouples can be connected to the DBK84 module. Channels 1 through 7 are
used for the first seven thermocouples, and channels 9 through 15 are used for the second set of seven.
Note: DBK84 has two on-board CJCs. They are measured on channels 0 and 8.
In addition to thermocouple measurements, each input channel can be configured for a fixed voltage gain
of 100. When in this mode, voltage can be measured in the range of ±100 mV, or ±50 mV, depending on
the type of Daq device being used.
Up to sixteen DBK84 modules can be attached to a single LogBook or Daq device, providing up to 224
temperature channels.
DBK Option Cards and Module 989494 DBK84 pg. 1
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - Overview

DBK84 14-Channel Thermocouple Module Overview …… 1 Hardware Setup …… 2 Module Connection …… 2 Open Thermocouple Detection …… 3 Module Configura

Page 2 - Hardware Setup

In comparison to typical DBK options, the DBK84 demands significant power from the system’s ±15 V power supplies. It is important that you calcul

Page 3 - Open Thermocouple Detection

through the use of shielded thermocouples and/or averaging. You can minimize the effect of noise by (1) using shielded thermocouples, (2) averagi

Page 4 - CAUTION

Module Configuration Up to sixteen DBK84 modules can be attached to a single LogBook or Daq device. Since multiple modules are connected via a paral

Page 5 - These jumpers do not

DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration Use of a DBK84 with a DaqBook/100 Series device, DaqBook/200 Series device

Page 6 - Software Setup

Software Setup Reference Notes: o DaqView users - Refer to Chapter 3, DBK Setup in DaqView. o LogView users - Refer to Chapter 4, DBK Setup in L

Page 7 - 989494 DBK84 pg. 7

DBK Option Cards and Module 989494 DBK84 pg. 7 Using a Temperature Calibrator The DBK84 thermocouple module provides accurate and repeatable t

Page 8 - DBK84 - Specifications

pg. 8, DBK84 989494 DBK Option Cards and Modules DBK84 - Specifications Name/Function: DBK84 – 14 Channel High-Accuracy Thermocouple Module S

Commentaires sur ces manuels

Pas de commentaire